Comparison of competing OPG & CBCT products from

2D

3D

NewTom GiANO HR DC

Technical Specifications

2D

Product

NewTom GiANO HR DC

Sensor type and dimensions

cmOS (CSI), 148×6 mm (151.2×6 mm); DC sensor option (direct conversion)

Kinematics (sensing method)

Multi–⁠motor kinematics with real–⁠time exposure and trajectory control according to patient morphology (SafeBeam) + multifocal technology (MultiPAN)

Exposure time (PAN and CEPH)

Pano: 5.7–⁠12.3 sec Ceph: 3.2–⁠9 sec

Dynamic Range

14 bit (16 bit for DC sensor)

Upgrade to Ceph

YES

Sensor for Ceph

Standalone or shared

TWAIN compatibility

YES (DICOM 3.0, TWAIN)

Function 

–⁠

Multi size option of Ceph (DxHcm)

–⁠

3D Upgradability 

–⁠

Sensor Pixel Size

–⁠

Focal Spot

0.5 mm

kV

2D: 60 kV – 85 kV 3D: 90 kV

mA

2–⁠16 mA

Dimension

–⁠

Weight

155 kg (without Ceph) 175 kg (with Ceph)

SW 

–⁠

More information

Can be equipped with Direct Conversion (DC) sensor, face-to-face positioning, incremental upgrade to 3D/CBCT, SafeBeam (automatic exposure settings based on pre-scan), MultiPan (5 or 11 for DC) of differently focused images, bite-wing mode, maximum ease of use.

2D

NewTom GiANO GO 2D

Technical Specifications

2D

Product

NewTom GiANO GO 2D

Sensor type and dimensions

cmOS (CSI), 6×151 mm

Kinematics (sensing method)

Multi–⁠motor kinematics with real–⁠time exposure and trajectory control according to patient morphology (SafeBeam) + multifocal technology (MultiPAN)

Exposure time (PAN and CEPH)

Pano: 6.8–⁠13.7 sec Ceph: 3.7–⁠9.9 sec

Dynamic Range

14 bit

Upgrade to Ceph

YES

Sensor for Ceph

Standalone

TWAIN compatibility

YES (DICOM 3.0, TWAIN)

Function 

–⁠

Multi size option of Ceph (DxHcm)

–⁠

3D Upgradability 

–⁠

Sensor Pixel Size

–⁠

Focal Spot

0.5 mm

kV

60–⁠90 kV

mA

4–⁠15 mA

Dimension

–⁠

Weight

2D version: 87 kg 3D/2D version: 99 kg CEPH option: +21 kg

SW 

–⁠

More information

SafeBeam (automatic adjustment of exposure parameters based on pre-scan), fast Eco scans, MultiPan (5 differently focused images per output), bite-wing mode, compact size, maximum ease of use.

2D

NewTom 7G

Technical Specifications

3D

Product

NewTom 7G

Functions (Device Type)

Dedicated CBCT

Voxel size

0.09–0.5 mm

Focal Spot Size

0.3–⁠0.6 mm

FOV size without stitching 

From 4×4 to 29×56 cm

FOV size with stitching

29×17 cm Maximum FOV: 29×56 cm

Scan Time / Exposure Time

Exposition: 1.4 – 8.8 sec Scan: 7.2 – 26 sec

Image data processing time

–⁠

Dynamic range

16 bit

Upgrade to Ceph

–⁠

Sensor for Ceph

–⁠

Patient position

Lying on a motorized table

Warranty

–⁠

Panoramic type

–⁠

kV

70–⁠120 kV

mA

Focal spot 0.3 mm → 5 –⁠54 mA Focal spot 0.6 mm → 55 –⁠120 mA

Dimension

–⁠

Weight

215 kg

SW 

–⁠

More information

The only whole-body CBCT in the world with applications in orthopedics, ENT, dentistry and veterinary medicine. Maximum FOV with minimum dose and ultimate resolution. 3D, 2D, CineX (X-ray video), Cine-scout (live Xray video projection)

3D

NewTom GiANO HR DC

Technical Specifications

3D

Product

NewTom GiANO HR DC

Functions (Device Type)

Hybrid system OPG/CBCT/CEPH

Voxel size

0.075 (0,068)–⁠0.3 mm; 0.075 mm resolution available for all FOVs; selected FOVs available from 0.068 mm

Focal Spot Size

0.5 mm

FOV size without stitching 

From 6×6 (4×4) to 10×8 cm (PRIME); 13×16 cm (ADVANCED); 16×18 cm (PROFESSIONAL)

FOV size with stitching

Without stitching

Scan Time / Exposure Time

2D: 5.7–12.3 sec 3D: 3.6–33.6 sec Exposition: 0.9–10.4 sec

Image data processing time

Real–⁠time image data processing

Dynamic range

3D: 16 bit 2D Standard: 14 bit 2D DC: 16 bit

Upgrade to Ceph

YES

Sensor for Ceph

Standalone or shared

Patient position

Standing, seated, wheelchair

Warranty

–⁠

Panoramic type

–⁠

kV

90 kV

mA

2–⁠16 mA

Dimension

–⁠

Weight

155 kg (without Ceph) 175 kg (with Ceph)

SW 

–⁠

More information

NewTom - World's first CBCT manufacturer; direct conversion (DC) sensor option; large FOV available at highest resolution; SafeBeam (automatic adjustment of 2D/3D exposure parameters based on pre-scan); MultiPan (5 or 11 for DC); bite-wing; compact size, maximum ease of use

3D

NewTom GO

Technical Specifications

3D

Product

NewTom GO

Functions (Device Type)

Hybrid system OPG/CBCT/CEPH

Voxel size

0.08–0.3 mm; 0.08 mm resolution available for all FOVs

Focal Spot Size

0.6 mm

FOV size without stitching 

6×6 mm to 10×10 cm; 8 sizes available

FOV size with stitching

Without stitching

Scan Time / Exposure Time

2D: 3.7–13.7 sec 3D: 6.4–16.8 sec

Image data processing time

Pano: Real Time CBCT: 15 sec

Dynamic range

3D: 16 bit 2D: 14 bit

Upgrade to Ceph

YES

Sensor for Ceph

Standalone or shared

Patient position

Standing, seated, wheelchair

Warranty

–⁠

Panoramic type

Amorphus Silicon (CSI)

kV

60–90 kV

mA

4–15 mA

Dimension

178.51 cm×112.49 cm×228.39 cm

Weight

113.85 kg (without Ceph) 134.71 kg (with Ceph)

SW 

NewTom NNT and App iPad NNTviewer

More information

NewTom - World's first CBCT manufacturer; large FOV available at highest resolution; SafeBeam (automatic adjustment of 2D/3D exposure parameters based on prescan); fast scans for 2D/3D; MultiPan (5 differently focused images per output); bite-wing; compact size, maximum ease of use. Auto Adaptive Pan / Auto Adaptive MAR

3D

NewTom VGi Evo

Technical Specifications

3D

Product

NewTom VGi Evo

Functions (Device Type)

Dedicated CBCT

Voxel size

0.1–0.3 mm

Focal Spot Size

0.3 mm

FOV size without stitching 

5×5 cm up to 24×19 cm

FOV size with stitching

Without stitching

Scan Time / Exposure Time

Exposition: 0,9–6 sec Scan: 15–25 sec

Image data processing time

Real–⁠time image data processing; Less than 1 minute

Dynamic range

16 bit

Upgrade to Ceph

NO

Sensor for Ceph

NO

Patient position

Standing, seated, wheelchair

Warranty

–⁠

Panoramic type

–⁠

kV

75–⁠110 kV

mA

1–⁠32 mA

Dimension

–⁠

Weight

Scan unit 377 kg Control box 95 kg

SW 

-

More information

NewTom - World's first CBCT manufacturer; High-end dedicated CBCT with highest quality imaging; SafeBeam (automatic adjustment of exposure parameters), rotating X-ray anode, CineX - serial X-ray imaging, video of joints, swallowing, ECO scans (ultra-low dose), Sharp2D OPG+CEPH in one scan

3D

The content of this document is intended for the Vatech authorized distributors in Central and Eastern Europe only.

It contains only publicly available information provided by the manufacturers themselves or their distributors on various websites or comparison sites.

But anyway, please do not share with anyone.

Get in touch

Social

© Copyright 2013 - 2024 MYDOKUMENT.COM - All rights reserved.